PhotoMetrics, Inc.
(714) 895-4465
Fax (714) 893-4682
When a sample undergoes a physical transformation such as a phase transition, more or less heat will need to flow to it than to the reference (typically an empty sample pan) to maintain both at the same temperature. Whether more or less heat must flow to the sample depends on whether the process is exothermic or endothermic. For example, as a solid sample melts to a liquid it will require more heat flowing to the sample to increase its temperature at the same rate as the reference. This is due to the absorption of heat by the sample as it undergoes the endothermic phase transition from solid to liquid. Likewise, as the sample undergoes exothermic processes (such as crystallization) less heat is required to raise the sample temperature. By observing the difference in heat flow between the sample and reference, differential scanning calorimeters are able to measure the amount of heat absorbed or released during such transitions. DSC may also be used to observe more subtle phase changes, such as glass transitions.
DSC is commonly used to measure a variety of
properties in both organic and inorganic materials, from metals and simple
compounds to polymers and pharmaceuticals. The properties measured include:
|
Home
|Techniques
| SEM
| FESEM
| EDS
| AFM
| WDS
| AUGER
| FTIR
| Literature Request |